DCM 200 Installation
Page 8
2.2 Mounting
1.1.
1.2. Fully Assembled System on Acrylic Backpanel
1.3.
1.4. Overall System Mounting
• Locate an area on a flat wall large enough to
accommodate the full size of the 22x22” system
panel with enough extra room to accommodate
flexible sample tubing. The controller is shipped
with a 3 ft. power cord already installed, for easy
connection to a close wall receptacle.
Alternatively, you may choose, or be required, to
install conduit into the bottom of the controller
enclosure. Refer to Section 4.4 for enclosure
opening sizes and locations.
• Refer to dimensional drawings in appendix section
for location and spacing of pre-drilled mounting
holes, measure and mark these locations on the
wall.
• Approximate weight of the assembled system with
water flowing may be in excess of 30 lbs., so take
care to choose adequately sized mounting hardware.
1.5. Individual Components
DCM 200 Controller Enclosure Mounting
•
Locate an area on a vertical wall where the
controller can me mounted at approximately eye
level and have enough clearance to allow the
DGMa modular sensor housing to be installed near
and below the controller. The location of the
sensor housing must have about 18” of clearance
above it to allow removal of the sensors, and about
12 inches below to access water sample petcock
and a sample cup.
NOTE: a common mistake made at installation, is
not allowing enough space above the sensor
housing to accommodate easy sensor maintenance.